MI 3321 MultiservicerXA
Technical specifications
19
3.3 Discharging time
Discharging time
Range
Resolution
Accuracy
0.0 s
… 9.9 s
0.1 s
(5 % of r 3 digit)
Peak voltage
Range
Resolution
Accuracy
0
V… 550 V
1 V
(5 % of r 5 digit)
2 wires system, triggered on DC voltage falling slope
Max. working voltage: ............... 550 V peak
Min. working voltage: ................ 85, 170 V peak
Threshold values: ..................... 1 s, 5 s
Threshold voltage level: ............ 60 V, 120 V
Input resistance: ....................... 48 M
Test terminals:
Discharging time
TP1 test terminal L↔N
3.4 Continuity
Continuity 10 A (PAT)
Range
Resolution
Accuracy
0.00
… 1.99
0.01
(5 % of r 3 digits)
Indicator only: 2.0
… 19.9
Continuity 10 A (other)
Range
Resolution
Accuracy
0.000
… 0.999
0.001
(5 % of r 6 digits)
1.00
… 1.99
0.01
(5 % of r 3 digits)
Indicator only: 2.0
… 19.9
Continuity 200 mA
Range
Resolution
Accuracy
0.00
… 1.99
0.01
(5 % of r 3 digits)
Indicator only: 2.0
… 19.9
Test currents: ............................ 10 A into 100 m
at rated mains supply voltage
200 mA into 2.00
Open circuit voltage: ................. <9 V AC
Test leads resistance compensation: .............................. yes,
Pass levels [
]: ......................... 0.01
… 0.09, 0.10 … 0.90, 1.00 … 9.00
Test duration [s]: ....................... 5, 10, 30, 60, 120, 180
Test method: ............................. 2-wire measurement, floating to earth
Test terminals:
Continuity (PAT)
S ↔ PE (mains test socket) ; (200 mA and 10 A)
S ↔ PE; (200 mA)
Continuity (other)
S/C1 ↔ C2