MI 3321 MultiservicerXA
General description
9
In PAT testing operating mode powerful functions for fast and efficient periodic testing
are included:
- Pre-programmed test sequences,
- Fast testing with barcode identification systems,
- Test data can be uploaded from PC,
- On site comparisons between old and new test results can be performed,
- Enables on site printing of test labels.
1.1 Warnings
In order to reach high level of operator safety while carrying out various measurements
using MultiservicerXA instrument, as well as to keep the test equipment undamaged, it
is necessary to consider the following general warnings:
Read this Instruction manual carefully, otherwise use of the instrument may
be dangerous for the operator, for the instrument or for the equipment under
test!
Warning on t
he instrument means »Read the Instruction manual with
special care to safety operation«. The symbol requires an action!
If the test equipment is used in manner not specified in this Instruction manual
the protection provided by the equipment may be impaired!
Do not use the instrument and accessories if any damage is noticed!
Consider all generally known precautions in order to avoid risk of electric
shock while dealing with hazardous voltages!
Do not use the instrument in supply systems with voltages higher than
300 V!
Use only standard or optional test accessories declared for this instrument
supplied by your distributor!
Regularly check the instrument and accessories for correct functioning to
avoid hazard that could occur from misleading results!
The protection class of the TP1 test terminal is CAT III / 300V. It means that
maximum allowed voltage between test terminals and ground is 300 V!
Correctly earthed mains sockets shall only be used to supply the instrument!
In case a fuse has blown follow the instructions in this Instruction manual to
replace it!
Instrument servicing and calibration is allowed to be carried out only by a
competent authorized person!
Hazardous voltages of up to 2200 V nominal are applied to the HV terminals
during the Withstanding test. It is advisable to keep the current trip out limit as
low as possible. Risk of electric shock!
It is advisable not to run tested devices with load currents above 13 A for more
than 15 minutes. Load currents higher than 13 A can result in excessive
temperatures of On/Off switch and fuse holders!
For working with the instrument use only original supplied power supply cord
with 1.5 mm2 wires cross-section (csa)!
All normal safety precautions have to be taken in order to avoid risk of electric
shock when working on electrical installations!