MI 3321 MultiservicerXA
Machine testing operating mode
119
Disconnect
test leads and proceed to the next measuring point. (no beep)
.............
.............
Connect
test leads to both ends of the last PE connection under test. The
instrument detects (double beep) the low resistance and starts last measurement.
After the timer elapse the instruments temporarily stores the last result. A sound
signal (short beeps) indicates that the last measurement is concluded and
temporarily stored. The results displayed in this phase will not be stored.
Disconnect
test leads (no beep).
Press
STOP
key to stop the measurement.
Press
SAVE
(F4) key for storing the results (optional). All temporarily stored results
are saved into the instruments memory.
Example of continuity measurement results
– normal view
Displayed results:
Main result ............. actual measured resistance / last measurement result
Rmax ..................... highest measured resistance
By pressing the
MORE
(F5) key all measured results can be seen.
Examples of continuity resistance measurement results
– all results view
Displayed results:
Main result ........................ actual measured resistance
Rmax ................................ highest measured resistance
Sub-results (1., 2., ...n) ..... temporarily stored individual results
By pressing the
BACK
(F5) key the instrument returns to normal view.
Notes
:
Consider displayed warnings before starting measurement!
STOP key can be pressed anytime during the measurement. All concluded
measurements are internally stored and can be saved to the instrument’s memory.
Optional external buzzer A1421 can be used in the Auto-continuity test.