MI 3321 MultiservicerXA
Single tests in PAT testing mode
50
This function is primarily intended for testing Class I DUTs.
Insulation menu
Test parameters for insulation resistance measurement
OUTPUT
Test
voltage
[250 V, 500 V]
LIMIT
Minimum resistance
[0.10 M
, 0.30 M
, 0.50 M
, 1.00 M
, 2.00
M
, 4.00 M
, 7.00 M
, 10.00 M
, none]
TIME
Measuring time
[5 s, 10 s, 30 s, 60 s, 120 s, none]
Test circuits for insulation resistance measurement
Measurement of insulation resistance of Class I DUT
Measurement of insulation resistance
of
fixed installed DUTs of Class I
Insulation resistance measurement procedure
Select the
Insulation
function.
Set test
parameters
.