Main menu and test modes
39
Keys in user edit menu:
Alphanumeric keys
Entering location name.
SAVE
(F1)
Confirms entry and returns to
User / device data menu
.
UNDO
(F2)
ESC
Discards modifications and returns to
User / device data menu
.
4.3.9
Password
In password menu two independent passwords can be set.
Password menu
PASSWORD HV-TEST protects unauthorised access to HV-TEST function. When
correct password is entered, HV-TEST functions are unlocked until the instrument is
switched off. This password cannot be disabled and can only be reset by
using “General
settings >> Original settings” option.
Note:
Password HV-TEST field is empty before first use of the instrument or after reset of
the instrument by applying original settings, but the instrument requires the
PASSWORD HV-TEST for operation with any of HV test function. For this reason,
the PASSWORD HV-TEST has to be set before first application of any HV test
function. The instrument keeps the password until it is changed or reset on applied
original settings.
PASSWORD GENERAL protects:
- Entering Edit user menu,
- Editing measurement parameters in single / autotest custom test mode,
- Deleting stored results,
- Entering Original settings menu.
In password protected actions, it is necessary to enter the PASSWORD GENERAL
before deleting or editing the protected data. The instrument requires a password and it
will not allow changes unless the correct password has been entered. This password is
disabled by default.