MI 3321 MultiservicerXA
Autotest sequences
81
7.2 Custom autotests
In autotest custom menu setting and editing user-defined autotest procedures is
allowed. Two sets (one for portable appliances and one for welding machines) of up to
50 custom autotest sequences can be pre-programmed in this autotest mode.
The most often used autotest sequences are added to the list by default.
The sequences can be also uploaded from the PC software PATLink PRO. Refer to
chapter
8.5
Data upload / download
for more information.
The pre-programmed sequences can be restored to default settings by selecting
Original settings
in
General settings menu
.
Select Custom Autotest in PAT testing main menu.
Custom autotest menu
– appliances
Custom autotest menu
– welding machines
Keys:
/
Selects the custom autotest.
VIEW
(F1)
Opens View menu for viewing details of selected test sequence
DELETE
(F3)
Removes selected test, see chapter
7.2.1 Deleting an existing custom
test sequence
.
APPL. /
WELDING
(F4)
Switches between the two sets of custom autotests (if applicable).
START
Starts the selected autotest. See chapter
7.5 Performing autotest
sequences
.
ESC
Returns to Main menu.
Notes:
For testing welding machines the 3 phase operation mode must be enabled and
the MultiservicerXA must be connected to a METREL 3-phase adapter (A1422).
See 3-phase adapter (A1322 / A1422) Instruction manual for more information.
If more than 50 autotests ar
e saved, »
Out of memory
« message is displayed.
HW 4.x or higher :
When PRCD test is enabled in autotest sequence then mains supply voltage is
applied on test socket during earth bond test (if selected in autotest sequence).