MI 3321 MultiservicerXA
Autotest sequences
87
7.3.2
Starting a project autotest
Select Project Autotest in PAT testing main menu.
Using the filters, search for the DUTs to be retested. After the DUT is found and
selected (in Project autotest result menu) begin retesting the DUT by pressing the
START key.
The autotest sequence will be the same as it was for the selected Project autotest.
Exactly the same DUT data (except time and user) will be offered to store in the
selected Project autotest.
When a new autotest is saved, it will get an »
I
« ticker. The original autotest will get a
»
I
« or »
PC
« ticker when the DUT is retested through the project autotest.
7.3.3
Comparison of results (evaluation of result trends)
If an autotest sequence was performed from the Project autotests menu, an additional
option TREND is offered in the
View result
menu after the autotest is finished.
View Project autotest results menu
In the Save results window, press the VIEW (F4) key the results to enter the view
results window. By pressing TREND (F4) key again a comparison between the old and
new test data will be performed. This is a very useful feature to evaluate result trends in
cases where results are relatively close to the limit.
Evaluation of test results
Compare results menu example
Meaning of trend symbols:
New result of particular test is better than last result.
Examples: New insulation resistance result is higher than old result.