MI 3321 MultiservicerXA
Technical specifications
29
Power supply: ........................... Class I
Pollution degree: ....................... 2
Degree of protection: ................ IP 50 (closed and locked cover)
.................................................. IP 20 main test socket
Case: ........................................ shock proof plastic / portable
Display: ..................................... 240*128 dots graphic matrix display with backlight
Memory: .................................... 6000 memory locations
Communication interface
RS232 interfaces: ..................... 1200 bps
… 115200 bps, 1 start bit, 8 data bits, 1 stop
bit
RS232 connectors: ................... 9-pin sub miniature type D, female
USB interface: .......................... 1200 bps
… 115200 bps
USB connector:......................... type B
Insulation:
Communication ports to PE: ..... 600 k
, 5 %
Dimensions (w×h×d): ................ 41.0 cm × 17.5 cm × 37.0 cm
Weight
(with standard accessories): ..... 8.4 kg
Reference conditions
Reference temperature range: .. 15
C
… 35
C
Reference humidity range: ........ 35 %
… 65 % RH
Operation conditions
Working temperature range: ..... 0
C
… 40
C
Maximum relative humidity: ...... 85 % RH (0
C
… 40
C), non-condensing
Storage conditions
Temperature range: .................. -10
C
… 60
C
Maximum relative humidity: ...... 90 % RH (-10
C
… 40
C)
80 % RH (40
C
… 60
C)
Accuracies apply for 1 year in reference conditions. Temperature coefficient outside
these limits is 0.2 % of measured value per
C plus 1 digit, otherwise noted.
Fuses
Test socket protection: .............. 2 x T16 A / 250 V, 1500 A, 6.3 mm
32 mm