MI 3321 MultiservicerXA
Autotest sequences
95
7.5.7
Differential leakage current
Measurement is described in chapter
6.2.6 Differential leakage current
.
If the differential leakage test fails or was skipped other tests will not be carried out
because of safety.
Keys
START
Starts the leakage current measurement.
Proceeds with the next leakage current measurement (in continuous
measurement mode only).
Proceed to the next autotest sequence measurement (in single
measurement mode only).
HELP (F2)
Displays leakage test help screens.
REPEAT (F3)
Repeats the leakage current measurement.
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
SKIP (F4)
Skips leakage current measurement.
END (F5)
Ends the autotest sequence.
7.5.8
Touch leakage current measurement
Measurement is described in chapter
6.2.7 Touch leakage current
.
If the touch leakage test fails or was skipped other tests will not be carried out because
of safety.
Keys
START
Starts the touch leakage current measurement.
Proceeds with the next touch leakage current measurement (in
continuous measurement mode only).
Proceed to the next autotest sequence measurement (in single
measurement mode only).
HELP (F2)
Displays touch leakage test help screens.
REPEAT (F3)
Repeats the touch leakage current measurement.
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
SKIP (F4)
Skips touch leakage current measurement.
END (F5)
Ends the autotest sequence.
7.5.9
Polarity test
Measurement is described in chapter
6.2.8 Polarity test
.
If the polarity test fails or was skipped other tests will not be carried out because of
safety.
Keys
START
Starts the polarity test.
Proceed to the next autotest sequence measurement.
HELP (F2)
Displays polarity test help screens.