MI 3321 MultiservicerXA
Machine testing operating mode
139
Test circuit for Discharging Time test
Discharging time test
Discharging time measurement procedure
Select the
Discharging Time
single test.
Set test
parameters
.
Connect
three wire test cable to the TP1 test port on the instrument.
Connect
L and N wires of the three wire test cable to terminals to be tested.
Press the
START
key to start the measurement.
Wait
until “Ready” is displayed
Disconnect
the machine. “Wait” is displayed until the residual voltage drops below
60 V. If the voltage at disconnection moment is too low to get an accurate result
“Repeat” is displayed. The measurement must be repeated
Press
SAVE
(F4) key for storing the result (optional).