MI 3321 MultiservicerXA
Technical specifications
23
-0.1
I
/ +0; I
= 0.5
I
N
Test current shape: ................... Sine-wave (AC), pulsed (A), smooth DC (B)*
DC offset for pulsed current: ..... 6 mA (typical)
RCD type: ................................. G (non-delayed), S (time-delayed)
Test current starting polarity: .... (+) or (-)
Voltage range: .......................... 40 V
… 264 V (45 Hz … 65 Hz)
Test terminals:
RCD testing
TP1 test socket
RCD test current selection (r.m.s. value calculated to 20 ms) according to EN 61009-1:
I
N × 1/2
I
N × 1
I
N × 2
I
N × 5
RCD I
I
N (mA) AC A
B*
AC
A
B*
AC
A
B
AC
A
B*
AC A B*
10
5
3.5
5
10
20
20
20
40
40
50
100 100
30
15
10.5 15
30
42
60
60
84
120 150
212 300
100
50
35
50
100
141
200
200
282
400 500
707 1000
300
150 105
150
300
424
600
600
848
n.a. 1500 n.a. n.a.
500
250 175
250
500
707
1000 1000 1410 n.a. 2500 n.a. n.a.
1000
500 350
500
1000 1410 n.a.
2000 n.a.
n.a. n.a.
n.a. n.a.
n.a.
n.a. ..................................................... not applicable
AC type .............................................. sine wave test current
A type…… ......................................... pulsed current
B type* ............................................... smooth DC current
Note:
All data (marked with “*”) regarding B type RCDs are valid for instrument HW version
3.0 and higher only.
3.13.2 Contact voltage RCD-Uc
Measuring range according to EN 61557-6 is 7.5 V
… 62.0 V for limit contact voltage
50 V.
Range
Resolution
Accuracy
0.0 V
… 19.9 V
0.1 V
(-
0 % / +15 %) of reading ± 10 digits
20.0 V
… 99.9 V
(-0 % / +15 %) of reading
The accuracy is valid if mains voltage is stable during the measurement and PE
terminal is free of interfering voltages.
Test current: ............................. max. 0.5
I
N
Limit contact voltage: ................ 50 V
Specified accuracy is valid for complete operating range.
3.13.3 Trip-out time
Complete measurement range corresponds to EN 61557-6 requirements.
Maximum measuring times set according to selected reference for RCD testing.
Range
Resolution
Accuracy
0 ms
… 40 ms
1 ms
1 ms