Single tests in PAT testing mode
59
Test circuits for touch leakage current measurement
Measurement of touch leakage current
Measurement of touch leakage current on a fixed installed DUT
Touch leakage current measurement procedure
Select the
Touch leakage
function.
Set test
parameters
.
Connect
S/C1 probe to accessible conductive parts of the DUT (see figures
above).
For
Portable appliance/device
:
Connect
device under test to the instrument.
For
fixed equipment
:
Power on
the fixed equipment;
Press the
START
key for measurement.
Examples of touch leakage current measurement results
Displayed results:
Main result ............. touch leakage current