MI 3321 MultiservicerXA
Autotest sequences
99
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
HELP (F2)
Displays the insulation resistance test help screens.
REPEAT (F3)
Repeats the insulation resistance measurement.
SKIP (F4)
Skips insulation resistance measurement.
END (F5)
Ends the autotest sequence.
7.6.4
Insulation resistance (welding circuit to protective circuit)
Measurement is described in chapter
6.3.3 Insulation resistance (welding circuit to
protective circuit)
. If this insulation test fails or was skipped other tests will not be carried
out because of safety.
Keys:
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
HELP (F2)
Displays the insulation resistance test help screens.
REPEAT (F3)
Repeats the insulation resistance measurement.
SKIP (F4)
Skips insulation resistance measurement.
END (F5)
Ends the autotest sequence.
7.6.3
Insulation resistance (supply circuit to welding circuit)
Measurement is described in chapter
6.3.4 Insulation resistance
(supply circuit to
welding circuit). If the insulation test fails or was skipped other tests will not be carried
out because of safety.
Keys:
START
Starts the insulation resistance measurement.
Proceeds with the next insulation resistance measurement
(in continuous mode).
Proceeds to the next autotest sequence measurement (in single
measurement mode only).
ENTER
Proceeds to the next autotest sequence measurement (in continuous
measurement mode only).
HELP (F2)
Displays the insulation resistance test help screens.
REPEAT (F3)
Repeats the insulation resistance measurement.
SKIP (F4)
Skips insulation resistance measurement.
END (F5)
Ends the autotest sequence.