MI 3321 MultiservicerXA
Single tests in PAT testing mode
70
Result screens:
Examples of Insulation LN-PE measurement results
Displayed results:
Main result ............. Insulation resistance LN-PE
Note
:
Consider displayed warnings before starting measurement!
6.3.3 Insulation resistance (welding circuit to protective circuit)
The insulation resistance test checks the resistance between the welding circuit
(outputs) and the protective circuit (protective earth) of the welding machine.
The instrument measures the insulation resistance between:
-
mains test socket‘s PE terminal and W1/ W2_TOUCH terminals on 3-phase
adapter (A1422).
Insulation W-PE menu
Test parameters for insulation resistance W-PE measurement
OUTPUT
Test
voltage
[500 V]
LIMIT
Minimum resistance
[2.50 M
, 5.00 M
, 10.00 M
, none]
TIME
Measuring time
[5 s, 10 s, 30 s, 60 s, 120 s, none]
Test circuit and measurement procedure for insulation resistance W-PE
measurement
For more information refer to chapter
Measurements according to IEC/ EN
60974-4
, paragraph
Insulation resistance (welding circuit to protective circuit)
in 3-phase adapter Instruction manual.