SETUP
ProMARK X
3-6
The currently selected mask angle is displayed on the Unit Ready display for
2 seconds when the unit is turned on. This will appear in the form MAxx on
the fourth line of the display, where xx is a number representing the mask
angle.
Access SETUP 4. Press the
RIGHT ARROW
to increase the mask angle
by increments of 1
°
until it gets to a mask angle of 20
°
. The next press
of the RIGHT ARROW sets the mask angle to 0
°
.
SETUP 5 — Data Sample Rate
. The data sample rate allows you to select
how quickly the unit saves information to an internal file or outputs it to a
PC or data logger. You may select separate data sample rates for the raw
data collected by the unit and for the position fixes it calculates.
When the receiver is on, the unit constantly collects raw data and calculates
position fixes. Both types of data are stored in a file or downloaded if the
dataport is on. In most instances, both types of data are not required. Since
you can set different sample rates for position fixes and raw data, you can
keep only the amount of each data type that you require for your application.
The data sample rate for either type of data can be set to any frequency from
0 seconds (no sampling) to every 999 seconds. Select 1 to save or output data
every second. The default data sample rate for both data types is 1.
Access SETUP 5. To select a sample rate for position fixes, press CLEAR,
then enter any number between 0 and 999 seconds. (Leading zeros are
not required.) Press
ENTER
.
If you want to change the raw data sample rate without altering position
sample rate, retype the same position rate and press
ENTER
. Now enter
the raw data sample rate.
To select a sample rate for raw data, enter any number between 0 and
999 seconds. Press
ENTER
.
NOTE
If you are logging data for carrier phase differential, you must
select 1 to sample raw data every second. This will allow you
to match data sets more easily.
SETUP 6 — Timer
. The Dif Timer allows you to set a starting time and
duration for a stationary differential session. (A session must be defined with
DIF 2 for the timer to work properly.) When the time is set to ON, any
stationary session you initiate will start at the designated time and for the
designated duration. This function remains on until turned off with
SETUP 6.