76
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
4
Math Waveforms
A scale unit of
U
(undefined) will be displayed for math functions when two source
channels are used and they are set to dissimilar units and the combination of units
cannot be resolved.
Math Operators
Math operators perform arithmetic operations (like add, subtract, or multiply) on
analog input channels.
•
•
Add or Subtract
When you select add or subtract, the
Source 1
and
Source 2
values are added or
subtracted point by point, and the result is displayed.
You can use subtract to make a differential measurement or to compare two
waveforms.
If your waveforms' DC offsets are larger than the dynamic range of the
oscilloscope's input channels you will need to use a differential probe instead.
Math function
Units
add or subtract
V or A
multiply
V
2
, A
2
, or W (Volt-Amp)
d/dt
V/s or A/s (V/second or A/second)
dt
Vs or As (V-seconds or A-seconds)
FFT
√
(square root)
V
1/2
, A
1/2
, or W
1/2
(Volt-Amp)
* When the FFT source is channel 1, 2, 3 or 4, FFT units will be displayed in dBV when channel units
is set to Volts and channel impedance is set to 1 M
Ω
. FFT units will be displayed in dBm when
channel units is set to Volts and channel impedance is set to 50
Ω
. FFT units will be displayed as dB
for all other FFT sources or when a source channel's units has been set to Amps.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...