266
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
17
Waveform Generator
To specify the expected output load
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Out Load
softkey and turn
the Entry knob to select:
•
50
Ω
•
High-Z
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
To use waveform generator logic presets
With logic level presets, you can easily set the output voltage to TTL, CMOS
(5.0V), CMOS (3.3V), CMOS (2.5V), or ECL compatible Low and High levels.
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Logic Presets
softkey.
4
In the Waveform Generator Logic Level Presets Menu, press one of the softkeys
to set the generated signal's Low and High voltages to logic compatible levels:
Waveform Type
Sync Signal Characteristics
All waveforms except
DC, Noise, and
Cardiac
The Sync signal is a TTL positive pulse that occurs when the waveform
rises above zero volts (or the DC offset value).
DC, Noise, and
Cardiac
N/A
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...