201
Keysight InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
13 Cursors
To make cursor measurements / 202
Cursor Examples / 205
Cursors are horizontal and vertical markers that indicate X-axis values and Y-axis
values on a selected waveform source. You can use cursors to make custom
voltage, time, phase, or ratio measurements on oscilloscope signals.
Cursor information is displayed in the right-side information area.
Cursors are not always limited to the visible display. If you set a cursor, then pan
and zoom the waveform until the cursor is off screen, its value will not be changed.
It will still be there when you return to its original location.
X Cursors
X cursors are vertical dashed lines that adjust horizontally and can be used to
measure time (s), frequency (1/s), phase (°), and ratio (%).
The X1 cursor is the short-dashed vertical line, and the X2 cursor is the
long-dashed vertical line.
When used with the FFT math function as a source, the X cursors indicate
frequency.
In XY horizontal mode, the X cursors display channel 1 values (Volts or Amps).
The X1 and X2 cursor values for the selected waveform source are displayed in the
softkey menu area.
The difference between X1 and X2 (
Δ
X) and 1/
Δ
X are displayed in the Cursors box
in the right-side information area.
Y Cursors
Y cursors are horizontal dashed lines that adjust vertically and can be used to
measure Volts or Amps, dependent on the channel
Probe Units
setting, or they can
measure ratios (%). When math functions are used as a source, the measurement
units correspond to that math function.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...