MIL-STD-1553/ARINC 429 Triggering and Serial Decode
27
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
401
MIL-STD-1553 Triggering
To set up the oscilloscope to capture a MIL-STD-1553 signal, see
To set up a MIL-STD-1553 trigger:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select the serial slot (Serial 1 or Serial 2) on which the MIL-STD-1553 signal is
being decoded.
3
Press the
Trigger
softkey; then, turn the Entry knob to select the trigger
condition:
•
Data Word Start
— triggers on the start of a Data word (at the end of a valid
Data Sync pulse).
•
Data Word Stop
— triggers on the end of a Data word.
•
Command/Status Word Start
— triggers on the start of Command/Status word
(at the end of a valid C/S Sync pulse).
•
Command/Status Word Stop
— triggers on the end of a Command/Status word.
•
Remote Terminal Address
— triggers if the RTA of the Command/Status word
matches the specified value.
When this option is selected, the
RTA
softkey becomes available and lets you
select the hex Remote Terminal Address value to trigger on. If you select
0xXX (don't cares), the oscilloscope will trigger on any RTA.
•
Remote Terminal A 11 Bits
— triggers if the RTA and the remaining 11
bits match the specified criteria.
When this option is selected, these softkeys become available:
•
The
RTA
softkey lets you select the hex Remote Terminal Address value.
•
The
Bit Time
softkey lets you select the bit time position.
•
The
0 1 X
softkey lets you set the bit time position value as a 1, 0, or X
(don't care).
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...