Trigger Mode/Coupling
11
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
181
The external trigger input impedance is 1M Ohm. This lets you use passive probes
for general-purpose measurements. The higher impedance minimizes the loading
effect of the oscilloscope on the device under test.
To set the EXT TRIG IN units and probe attenuation:
1
Press the
[Mode/Coupling]
key in the Trigger section of the front panel.
2
In the Trigger Mode and Coupling Menu, press the
External
softkey.
3
In the External Trigger Menu, press the
Units
softkey to select between:
•
Volts
— for a voltage probe.
•
Amps
— for a current probe.
Measurement results, channel sensitivity, and the trigger level will reflect the
measurement units you have selected.
4
Press the
Probe
softkey; then, turn the entry knob to specify the probe
attenuation.
The attenuation factor can be set from 0.001:1 to 10000:1 in a 1-2-5 sequence.
The probe attenuation factor must be set properly for measurements to be
made correctly.
CAUTION
Maximum voltage at oscilloscope external trigger input
300 Vrms, 400 Vpk
1 M ohm input: For steady-state sinusoidal waveforms derate at 20 dB/decade above
57 kHz to a minimum of 5 Vpk
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...