68
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
3
Vertical Controls
When
Fine
is turned off, turning the volts/division knob changes the channel
sensitivity in a 1-2-5 step sequence.
To invert a waveform
1
Press the desired channel key.
2
In the Channel Menu, press the
Invert
softkey to invert the selected channel.
When
Invert
is selected, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed. However, when using basic triggers, the
oscilloscope attempts to maintain the same trigger point by changing trigger
settings.
Inverting a channel also changes the result of any math function selected in the
Waveform Math Menu or any measurement.
Setting Analog Channel Probe Options
1
Press the probe's associated channel key.
2
In the Channel Menu, press the
Probe
softkey to display the Channel Probe
Menu.
This menu lets you select additional probe parameters such as attenuation
factor and units of measurement for the connected probe.
The Channel Probe Menu changes depending on the type of probe connected.
For passive probes (such as the N2841A, N2842A, N2843A, N2862A/B,
N2863A/B, N2889A, N2890A, 10073C, 10074C, or 1165A probes), the
Probe
Check
softkey appears; it guides you through the process of compensating
probes.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...