150
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
• When the time range (
><
) qualifier is selected, the Entry knob sets the upper
time range value.
Pulse width
trigger > qualifier
time set softkey
• When the greater than (
>
) qualifier is selected, the Entry knob sets the
oscilloscope to trigger on a pulse width greater than the time value displayed
on the softkey.
• When the time range (
><
) qualifier is selected, the Entry knob sets the lower
time range value.
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified pattern.
This pattern is a logical AND combination of the channels. Each channel can have
a value of 0 (low), 1 (high), or don't care (X). A rising or falling edge can be
specified for one channel included in the pattern. You can also trigger on a hex
bus value as described on
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select
Pattern
.
3
Press the
Qualifier
softkey; then, turn the Entry knob to select from the pattern
duration qualifier options:
•
Entered
– when the pattern is entered.
•
<
(Less Than) – when the pattern is present for less than a time value.
•
>
(Greater Than) – when the pattern is present for greater than a time value.
The trigger occurs when the pattern exits (not when the > softkey time value
is exceeded).
•
Timeout
– when the pattern is present for greater than a time value. In this
case, the trigger occurs when the > softkey time value is exceeded (not when
the pattern exits).
•
>< (In Range)
– when the pattern is present for a time within a range of values.
•
<> (Out of Range)
– when the pattern is present for a time outside of range of
values.
Pattern durations are evaluated using a timer. The timer starts on the last edge
that makes the pattern (logical AND) true. Except when the
Timeout
qualifier is
selected, the trigger occurs on the first edge that makes the pattern false,
provided the time qualifier criteria has been met.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...