382
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
25
I2C/SPI Triggering and Serial Decode
In the SPI Chip Select Menu:
a
Press the
Frame by
softkey to select a framing signal that the oscilloscope will
use for determining which clock edge is the first clock edge in the serial
stream.
You can set the oscilloscope to trigger during a high chip select (
CS
), a low
chip select (
~CS
), or after a
Timeout
period during which the clock signal has
been idle.
•
If the framing signal is set to
CS
(or
~CS
), the first clock edge as defined,
rising or falling, seen after the
CS
(or
~CS
) signal transitions from low to
high (or high to low) is the first clock in the serial stream.
Chip Select
— Press the
CS
or
~CS
softkey; then, turn the Entry knob to
select the channel that is connected to the SPI frame line. The label (~CS
or CS) for the source channel is automatically set. The data pattern and
the clock transition must occur during the time when the framing signal is
valid. The framing signal must be valid for the entire data pattern.
•
If the framing signal is set to
Timeout
, the oscilloscope generates it's own
internal framing signal after it sees inactivity on the serial clock line.
Clock Timeout
— Select
Clock Timeout
in the
Frame by
softkey, then select
the
Timeout
softkey and turn the Entry knob to set the minimum time that
the Clock signal must be idle (not transitioning) before the oscilloscope
will search for the Data pattern on which to trigger.
The Timeout value can be set anywhere from 100 ns to 10 s.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...