198
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
The
Analyze Segments
softkey appears when the acquisition is stopped and the
segmented memory feature is on or when the serial Lister is enabled.
You can also turn on infinite persistence (in the Display Menu) and press the
Analyze Segments
softkey to create an infinite persistence display.
Segmented Memory Re-Arm Time
After each segment fills, the oscilloscope re-arms and is ready to trigger in about
1 µs.
Remember though, for example: if the horizontal time per division control is set to
5 µs/div, and the Time Reference is set to
Center
, it will take at least 50 µs to fill all
ten divisions and re-arm. (That is 25 µs to capture pre-trigger data and 25 µs to
capture post-trigger data.)
Saving Data from Segmented Memory
You can save either the currently displayed segment (
Save Segment - Current
), or all
segments (
Save Segment - All
) in the following data formats: CSV, ASCII XY, and
BIN.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...