Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
15
Mask Testing
Run Until
The Run Until softkey lets you specify a condition on which to terminate testing.
• Forever
— The oscilloscope runs continuously. However, if an error occurs the
action specified using the
On Error
softkey will occur.
• Minimum # of Tests
— Choose this option and then use the
# of Tests
softkey
to select the number of times the oscilloscope will trigger, display the
waveform(s), and compare them to the mask. The oscilloscope will stop after
the specified number of tests have been completed. The specified minimum
number of tests may be exceeded. If an error occurs the action specified
using the
On Error
softkey will occur. The actual number of tests completed
is displayed above the softkeys.
• Minimum Time
— Choose this option and then use the
Test Time
softkey to
select how long the oscilloscope will run. When the selected time has passed
the oscilloscope will stop. The specified time may be exceeded. If an error
occurs the action specified using the
On Error
softkey will occur. The actual
test time is displayed above the softkeys.
• Minimum Sigma
— Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the oscilloscope
will perform the action specified by the
On Error
softkey.) Note that this is a
test sigma (the max achievable process sigma, assuming no defects, for a
certain number of tested waveforms) as opposed to a process sigma (which is
tied to the amount of failures per test). The sigma value may exceed the
selected value when a small sigma value is chosen. The actual sigma is
displayed.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...