348
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
23
CAN/LIN Triggering and Serial Decode
3
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
SOF - Start of Frame
— The oscilloscope triggers at the start of a frame.
•
Remote Frame ID (RTR)
— The oscilloscope triggers on remote frames with the
specified ID. Press the
Bits
softkey to select the ID.
•
Data Frame ID (~RTR)
— The oscilloscope will trigger on data frames matching
the specified ID. Press the
Bits
softkey to select the ID.
•
Remote or Data Frame ID
— The oscilloscope will trigger on remote or data
frames matching the specified ID. Press the
Bits
softkey to select the ID.
•
Data Frame ID and Data
— The oscilloscope will trigger on data frames
matching the specified ID and data. Press the
Bits
softkey to select the ID
and set up the number of data bytes and values.
•
Error Frame
— The oscilloscope will trigger on CAN active error frames.
•
All Errors
— The oscilloscope will trigger when any form error or active error is
encountered.
•
Acknowledge Error
— The oscilloscope will trigger when the acknowledge bit is
recessive (high).
•
Overload Frame
— The oscilloscope will trigger on CAN overload frames.
4
If you select a condition that lets you trigger on ID or data values, use the
Bits
softkey and the CAN Bits Menu to specify those values.
For details about using the CAN Bits Menu softkeys, press and hold the softkey
in question to display the built-in help.
You can use the
Zoom
mode for easier navigation of the decoded data.
NOTE
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the
oscilloscope is AutoTriggering. Press the
[Mode/Coupling]
key, then press the
Mode
softkey
to set the trigger mode from
Auto
to
Normal
.
NOTE
To display CAN serial decode, see
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...