I2S Triggering and Serial Decode
26
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
393
3
Press the
Trigger Setup
softkey to open the I
2
S Trigger Setup Menu.
4
Press the
Audio
softkey; then, turn the Entry knob to choose to trigger on
Left
channel events,
Right
channel events, or events that occur on
Either
channel.
5
Press the
Trigger
softkey and choose a qualifier:
•
Equal
— triggers on the specified audio channel's data word when it equals
the specified word.
•
Not equal
— triggers on any word other than the specified word.
•
Less than
— triggers when the channel's data word is less than the specified
value.
•
Greater than
— triggers when the channel's data word is greater than the
specified value.
•
In Range
— enter upper and lower values to specify the range in which to
trigger.
•
Out of Range
— enter upper and lower values to specify range in which trigger
will not occur.
•
Increasing value
— triggers when the data value is increasing over time and
the specified value is met or exceeded. Set
Trigger >=
to the data value that
must be reached. Set
Armed <=
to the value to which the data must fall
before the trigger circuit is re-armed (ready to trigger again). These settings
are made in the current menu when
Base
is
Decimal
or in the Bits submenu
when the
Base
is
Binary
. The Armed control reduces triggers due to noise.
This trigger condition is best understood when the digital data transferred
over the I2S bus is considered in terms of representing an analog waveform.
The figure below shows a plot of sample data transmitted over an I2S bus for
one channel. In this example, the oscilloscope will trigger at the 2 points
shown since there are two instances in which the data increases from a value
below (or at) the "Armed" value to a value greater than (or equal to) the
specified "Trigger" value.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...