Math Waveforms
4
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
75
To perform transforms or filters on an arithmetic operation
To perform a transform function (see
on page 78) or filter (see
on page 94) on the add, subtract, or multiply arithmetic operations:
1
Press the
Function
softkey and select
g(t): Internal
.
2
Use the
Operator
,
Source 1
, and
Source 2
softkeys to set up an arithmetic
operation.
3
Press the
Function
softkey and select
f(t): Displayed
.
4
Use the
Operator
softkey to select a transform function or filter.
5
Press the
Source 1
softkey and select
g(t)
as the source. Note that
g(t)
is only
available when you select a transform function in the previous step.
To adjust the math waveform scale and offset
1
Make sure the multiplexed scale and position knobs to the right of the
[Math]
key are selected for the math waveform.
If the arrow to the left of the
[Math]
key is not illuminated, press the key.
2
Use the multiplexed scale and position knobs just to the right of the
[Math]
key
to re-size and re-position the math waveform.
See Also
•
Units for Math Waveforms
Units for each input channel can be set to Volts or Amps using the
Units
softkey in
the channel's Probe Menu. Units for math function waveforms are:
NOTE
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is
automatically scaled for optimum vertical scale and offset. If you manually set scale and offset
for a function, select a new function, then select the original function, the original function will
be automatically rescaled.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...