Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
14
Measurements
5
Press the
Middle
softkey; then, turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper thresholds.
The default middle threshold is 50% or 1.20 V.
•
If threshold
Type
is set to
%
, the middle threshold value can be set from 6% to
94%.
6
Press the
Upper
softkey; then, turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will automatically
decrease the middle value to be less than the upper value. The default upper
threshold is 90% or 1.50 V.
•
If threshold
Type
is set to
%
, the upper threshold value can be set from 7% to
95%.
Measurement Window with Zoom Display
When the zoomed time base is displayed, you can choose whether measurements
are made in the Main window portion of the display or the Zoom window portion of
the display.
1
Press the
[Meas]
key.
2
In the Measurement Menu, press the
Settings
softkey.
3
In the Measurement Settings Menu, press the
Meas Window
softkey; then, turn
the Entry knob to select from:
•
Auto Select
— The measurement is attempted in the lower, Zoom window; if it
cannot be made there, the upper, Main window is used.
•
Main
— The measurement window is the upper, Main window.
•
Zoom
— The measurement window is the lower, Zoom window.
Measurement Statistics
Press the
[Meas]
key to enter the Measurement Menu. By default, statistics are
displayed, and frequency and voltage are measured on channel 1.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...