196
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
When using segmented memory, use the Analyze Segments feature (see
"Measurements, Statistics, and Infinite Persistence with Segmented Memory"
page 197) to show infinite persistence across all acquired segments. See also
on page 131 for details.
To acquire to
segmented
memory
1
Set up a trigger condition. (See
, “Triggers,” starting on page 141
for details.)
2
Press the
[Acquire]
key in the Waveform section of the front panel.
3
Press the
Segmented
softkey.
4
In the Segmented Memory Menu, press the
Segmented
softkey to enable
segmented memory acquisitions.
5
Press the
# of Segs
softkey and turn the Entry knob to select the number of
segments into which you would like to divide the oscilloscope's memory.
Memory can be divided into as few as two segments and as many as 500
segments with the standard 2M points of memory or 1000 segments with the
upgrade to 4M points memory.
6
Press the
[Run]
or
[Single]
key.
The oscilloscope runs and fills a memory segment for each trigger event. When
the oscilloscope is busy acquiring multiple segments, the progress is displayed in
the upper right area of the display. The oscilloscope continues to trigger until
memory is filled, then the oscilloscope stops.
If the signal you are measuring has more than about 1 s of inactivity, consider
selecting the
Normal
trigger mode to prevent AutoTriggering. See
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...