Waveform Generator
17
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
259
Sine
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
/
High-Level
, and
Offset
/
Low-Level
softkeys to set the sine signal parameters.
100 mHz to
20 MHz
20 mVpp to
5 Vpp
±2.50 V
Square
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Duty Cycle
softkeys to set the square wave signal
parameters.
The duty cycle can be adjusted from 20% to 80%.
100 mHz to
10 MHz
20 mVpp to
5 Vpp
±2.50 V
Ramp
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Symmetry
softkeys to set the ramp signal
parameters.
Symmetry represents the amount of time per cycle
that the ramp waveform is rising and can be
adjusted from 0% to 100%.
100 mHz to
200 kHz
20 mVpp to
5 Vpp
±2.50 V
Pulse
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
/
High-Level
,
Offset
/
Low-Level
, and
Width/Width Fine
softkeys to set the pulse signal
parameters.
The pulse width can be adjusted from 20 ns to the
period minus 20 ns.
100 mHz to
10 MHz.
20 mVpp to
5 Vpp
±2.50 V
DC
Use the
Offset
softkey to set the DC level.
n/a
n/a
±2.50 V
Noise
Use the
Amplitude
/
High-Level
and
Offset
/
Low-Level
to set the noise signal parameters.
n/a
20 mVpp to
5 Vpp
±2.50 V
Sine Cardinal
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
, and
Offset
softkeys to set the sinc
signal parameters.
100 mHz to
1 MHz
20 mVpp to
5 Vpp
±1.25 V
Exponential
Rise
Use the
Frequency/Frequency Fine/Period/Period
Fine
,
Amplitude
/
High-Level
, and
Offset
/
Low-Level
softkeys to set the exponential rise signal
parameters.
100 mHz to
5 MHz
20 mVpp to
5 Vpp
±2.50 V
Waveform
Type
Characteristics
Frequency
Range
Max.
Amplitude
(High-Z)
1
Offset
(High-Z)
1
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...