Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
28
UART/RS232 Triggering and Serial Decode
6
For both the Rx and Tx signals:
a
Connect an oscilloscope channel to the signal in the device under test.
b
Press the
Rx
or
Tx
softkey; then, turn the Entry knob to select the channel for
the signal.
c
Press the corresponding
Threshold
softkey; then, turn the Entry knob to select
the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become the
trigger level when the trigger type is set to the selected serial decode slot.
The RX and TX labels for the source channels are automatically set.
7
Press the
Back/Up key to return to the Serial Decode Menu.
8
Press the
Bus Config
softkey to open the UART/RS232 Bus Configuration Menu.
Set the following parameters.
a #Bits
— Set the number of bits in the UART/RS232 words to match your
device under test (selectable from 5-9 bits).
b Parity
— Choose odd, even, or none, based on your device under test.
c Baud
— Press the
Baud Rate
softkey, then press the
Baud
softkey and select a
baud rate to match the signal in your device under test. If the desired baud
rate is not listed, select
User Defined
on the Baud softkey; then, select the
desired baud rate using the
User Baud
softkey.
You can set the UART baud rate from 1.2 kb/s to 8.0000 Mb/s in increments
of 100 b/s.
d Polarity
— Select idle low or idle high to match your device under test's state
when at idle. For RS232 select idle low.
e Bit Order
— Select whether the most significant bit (MSB) or the least
significant bit (LSB) is presented after the start bit in the signal from your
device under test. For RS232 select LSB.
Back
NOTE
In the serial decode display, the most significant bit is always displayed on the left regardless
of how Bit Order is set.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...