Vertical Controls
3
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
67
When AC coupling is selected (see
page 65) or excessive voltage is applied to the input, the oscilloscope
automatically switches to
1M Ohm
mode to prevent possible damage.
•
1M Ohm
— is for use with many passive probes and for general-purpose
measurements. The higher impedance minimizes the loading effect of the
oscilloscope on the device under test.
This impedance matching gives you the most accurate measurements because
reflections are minimized along the signal path.
See Also
• For more information on probing, visit:
www.keysight.com/find/scope_probes
• Information about selecting a probe can be found in document number
Keysight Oscilloscope Probes and Accessories Selection Guide
(part number
5989-6162EN), available at
.
To specify bandwidth limiting
1
Press the desired channel key.
2
In the Channel Menu, press the
BW Limit
softkey to enable or disable bandwidth
limiting.
When bandwidth limit is on, the maximum bandwidth for the channel is
approximately 20 MHz. For waveforms with frequencies below this, turning
bandwidth limit on removes unwanted high frequency noise from the waveform.
The bandwidth limit also limits the trigger signal path of any channel that has
BW
Limit
turned on.
To change the vertical scale knob's coarse/fine adjustment setting
1
Push the channel's vertical scale knob (or press the channel key and then the
Fine
softkey in the Channel Menu) to toggle between fine and coarse
adjustment of the vertical scale.
When
Fine
adjustment is selected, you can change the channel's vertical
sensitivity in smaller increments. The channel sensitivity remains fully calibrated
when
Fine
is on.
The vertical scale value is displayed in the status line at the top of the display.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...