300
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
20
Utility Settings
•
Preserve
— to make the oscilloscope remain in the acquisition mode you
have chosen when the
[AutoScale]
key is pressed.
Setting the Oscilloscope's Clock
The Clock Menu lets you set the current date and time of day (24-hour format).
This time/date stamp will appear on hardcopy prints and directory information on
the USB mass storage device.
To set the date and time, or to view the current date and time:
1
Press
[Utility] > Options > Clock
.
2
Press the
Year
,
Month
,
Day
,
Hour
or
Minute
softkey; then, rotate the Entry knob to
set to the desired number.
The hours are shown in the 24-hour format. So 1:00 PM is hour 13.
The real-time clock only allows selection of valid dates. If a day is selected and the
month or year is changed so the day is invalid, the day is automatically adjusted.
Setting the Rear Panel TRIG OUT Source
You can choose the source of the TRIG OUT connector on the rear panel of the
oscilloscope:
1
Press
[Utility] > Options > Rear Panel
.
If the
Rear Panel
softkey does not appear, it means neither mask testing nor the
waveform generator is licensed, and oscilloscope triggers are sent to TRIG OUT
by default.
2
In the Rear Panel Menu, press
Trig Out
; then, turn the Entry knob to select from:
•
Triggers
— Each time the oscilloscope triggers, a rising edge occurs on
TRIG OUT. The rising edge is delayed 30 ns from the oscilloscope's trigger
point. The output level is 0-5 V into an open circuit, and 0-2.5 V into 50
Ω
.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...