Waveform Generator
17
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
267
To add noise to the waveform generator output
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Add Noise
softkey and turn
the Entry knob to select the amount of white noise to add to the waveform
generator output.
Note that adding noise affects edge triggering on the waveform generator source
(see
on page 144) as well as the waveform generator sync pulse
output signal (which can be sent to TRIG OUT, see
on page 300). This is because the trigger comparator is located after
the noise source.
To add modulation to the waveform generator output
Modulation is where an original carrier signal is modified according to the
amplitude of a second modulating signal. The modulation type (AM, FM, or FSK)
specifies how the carrier signal is modified.
To enable and set up modulation for the waveform generator output:
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
Softkey (logic levels)
Low level
High level, 50 ohm
expected output load
High level, high-Z
expected output load
TTL
0 V
+2.5 V (TTL compatible)
+5 V
CMOS (5.0V)
0 V
Not Available
+5 V
CMOS (3.3V)
0 V
+2.5 V (CMOS
compatible)
+3.3 V
CMOS (2.5V)
0 V
+2.5 V
+2.5 V
ECL
-1.7 V
-0.8 V (ECL compatible)
-0.9 V
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...