152
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10
Triggers
•
X
sets the pattern to don't care on the selected channel. Any channel set to
don't care is ignored and is not used as part of the pattern. However, if all
channels in the pattern are set to don't care, the oscilloscope will not trigger.
•
The rising edge ( ) or falling edge ( ) softkey sets the pattern to an edge
on the selected channel. Only one rising or falling edge can be specified in
the pattern. When an edge is specified, the oscilloscope will trigger at the
edge specified if the pattern set for the other channels is true.
If no edge is specified, the oscilloscope will trigger on the last edge that
makes the pattern true.
Hex Bus Pattern Trigger
You can specify a bus value on which to trigger. To do this, first define the bus. See
"To display digital channels as a bus"
on page 114 for details. You can trigger on
a bus value whether you are displaying the bus or not.
To trigger on a bus value:
1
Press the
[Pattern]
key on the front panel.
2
Press the
Channel
softkey and rotate the Entry knob to select
Bus1
or
Bus2
.
3
Press the
Digit
softkey and rotate the Entry knob to select a digit of the selected
bus.
4
Press the
Hex
softkey and rotate the Entry knob to select a value for the digit.
5
You can use the
Set all Digits
softkey to set all digits to a particular value.
When a hex bus digit contains one or more don't care (X) bits and one or more bit
with a value or 0 or 1, the "$" sign will be displayed for the digit.
NOTE
Specifying an Edge in a Pattern
You are allowed to specify only one rising or falling edge term in the pattern. If you define an
edge term, then select a different channel in the pattern and define another edge term, the
previous edge definition is changed to a don't care.
NOTE
If a digit is made up of less than four bits, then the value of the digit will be limited to the
value that can be created by the selected bits.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...