Chapter 10
291
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:TEST:HANDler:STATus:RDYTrig
Syntax
:TEST:HANDler:STATus:RDYTrig {ON|OFF|1|0}
Description
This command sets the Handler Ready for Trigger signal for test purpose and can be used
to check/troubleshoot the Test pin signal of Handler interface. (No Query)
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Related commands
:TEST:HANDler:STATus:ALARm on page 286
:TEST:HANDler:STATus:INDex on page 288
:TEST:HANDler:STATus:NC on page 289
:TEST:HANDler:STATus:OVLD on page 290
Equivalent key
sequence
No equivalent key is available on the front panel.
:TEST:HANDler:TRIGger?
Syntax
:TEST:HANDler:TRIGger?
Description
This command gets the Handler Trigger signal for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (Query Only)
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Query response
{1|0}<newline><^END>
Related commands
:TEST:REAR:TRIGger? on page 292
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON
Sets Handler Ready for Trigger signal status to ON.
OFF
Sets Handler Ready for Trigger signal status to OFF.