Chapter 10
237
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
• When specifying STAN3 as parameter 1:
In any case, if the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if the upper
limit of the range is exceeded) is set.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
[:SENSe]:CORRection:OPEN[:STATe] on page 244
[:SENSe]:CORRection:SHORt[:STATe] on page 245
[:SENSe]:CORRection:LOAD[:STATe] on page 238
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
[:SENSe]:CORRection:CKIT:STAN1:FORMat on page 229
[:SENSe]:CORRection:CKIT:STAN2:FORMat on page 230
[:SENSe]:CORRection:CKIT:STAN3:FORMat on page 232
:FORMat[:DATA] on page 222
Equivalent key
sequence
] - CORRECTION - SHORT- A|B
[
Meas Setup
] - CORRECTION - LOAD- A|B
Parameter 2: <numeric 1>
Parameter 3: <numeric 2>
Description
Value of the primary parameter
*1
.
Value of the secondary
parameter
*1
.
Range
-999.999 to 999.999
-99.9999E9 to 99.9999E9
Initial value 100E-9
0
Unit
F (farad)
Depends on the type of parameter.
*1.Use the
[:SENSe]:CORRection:CKIT:STAN3:FORMat
command to specify the type
of parameter.