78
Chapter 7
Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function
Setting Up Comparator Function
Turning ON/OFF comparator function
To enable/disable the comparator function, use the following command:
•
:CALCulate1:COMParator[:STATe] on page 187
Setting up limit range
Clearing (resetting) limit range
You can clear the ON/OFF state and the lower and upper limit values of all limit ranges
(BIN1 to BIN9 and the secondary parameter limit range) and the ON/OFF state of AUX
BIN to recover the factory-shipped default values (refer to
Operation Manual, Appendix C
“ Initial Setting”
).
To clear the limit ranges, use the following command.
•
:CALCulate1:COMParator:CLEar on page 176
Selecting limit range designation method
You can select the designation method of the primary parameter limit ranges (BIN1 to
BIN9) from the following three modes:
NOTE
Only the absolute mode can be used as the designation method of the secondary parameter
limit range.
To select the designation method, use the following command.
•
:CALCulate1:COMParator:MODE on page 181
To set up the reference value for the absolute tolerance mode or percent tolerance mode,
use the following command.
•
:CALCulate1:COMParator:PRIMary:NOMinal on page 184
Mode
Description
Absolute mode
Designation using the absolute value
Absolute tolerance mode
Designation using the
relative value (deviation
from the reference value)
Designation using the absolute
value
*1
Percent tolerance mode
Designation using the percentage of
the reference value
*2
*1.Boundary value - reference value
*2.((Boundary value - reference value)/reference value)
∞
100