Chapter 10
289
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:TEST:HANDler:STATus:NC
Syntax
:TEST:HANDler:STATus:NC {ON|OFF|1|0}
Description
This command sets the Handler No Contact/Low C Reject signal for test purpose and can
be used to check/troubleshoot the Test pin signal of Handler interface. (No Query)
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Related commands
:TEST:HANDler:STATus:ALARm on page 286
:TEST:HANDler:STATus:INDex on page 288
:TEST:HANDler:STATus:OVLD on page 290
:TEST:HANDler:STATus:RDYTrig on page 291
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON
Sets No Contact/Low C Reject status to ON.
OFF
Sets No Contact/Low C Reject status to OFF.