232
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:CKIT:STAN3:FORMat
Syntax
[:SENSe]:CORRection:CKIT:STAN3:FORMat
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}
[:SENSe]:CORRection:CKIT:STAN3:FORMat?
Description
Sets the types of parameters used to define the standard for LOAD correction.
Parameters
For details on Cp and Cs, refer to
:CALCulate1:FORMat on page 188
. For details on D, Q,
Rp, and Rs, refer to
:CALCulate2:FORMat on page 191
Query response
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}<newline><^END>
Related commands
[:SENSe]:CORRection:CKIT:STAN3[:DATA] on page 231
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - LOAD- Cp
{CPD|CPQ|CPG|CPRP}/ Cs{CSD|CSQ|CSRS}
Description
CPD
(initial value)
Specifies Cp as the primary parameter, D as the secondary parameter.
CPQ
Specifies Cp as the primary parameter, Q as the secondary parameter.
CPG
Specifies Cp as the primary parameter, G as the secondary parameter.
CPRP
Specifies Cp as the primary parameter, Rp as the secondary parameter.
CSD
Specifies Cs as the primary parameter, D as the secondary parameter.
CSQ
Specifies Cs as the primary parameter, Q as the secondary parameter.
CSRS
Specifies Cs as the primary parameter, Rs as the secondary parameter.