Chapter 10
231
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe]:CORRection:CKIT:STAN3[:DATA]
Syntax
[:SENSe]:CORRection:CKIT:STAN3[:DATA] <numeric 1>,<numeric 2>
[:SENSe]:CORRection:CKIT:STAN3[:DATA]?
Description
Defines the values of the LOAD correction standard for the parameters you specify by
using the
[:SENSe]:CORRection:CKIT:STAN3:FORMat
These are set as the standard values for the measurement frequency when executing the
command (set with the
command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
command) with the channel-by-channel
definition of the standard values enabled (set to ON with the
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]
command), the standard values
for the selected channel (selected with the
[:SENSe]:CORRection:MULTiple:CHANnel
command) are set at execution of the command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
[:SENSe]:CORRection:LOAD[:STATe] on page 238
[:SENSe]:CORRection:CKIT:STAN3:FORMat on page 232
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] on page 240
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - REF- A|B
<numeric 1>
<numeric 2>
Description
Value of the primary
parameter.
Value of the secondary
parameter.
Range
-999.999 to 999.999
-99.9999E9 to 99.9999E9
Initial value
100E-9
0
Unit
F (farad)
Depends on the type of
parameter.