Chapter 10
249
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe][:FIMPedance]:APERture:TIME
Syntax
[:SENSe][:FIMPedance]:APERture:TIME {1|2|4|6|8}
[:SENSe][:FIMPedance]:APERture:TIME?
Description
Specifies the measurement time.
For information on the specific measurement time of each mode, see “Specification and
Supplemental Performance Characteristics” in the
Operation Manual
.
Parameters
Query response
{1|2|4|6|8}<newline><^END>
Equivalent key
sequence
[
Meas Setup
] - MEAS TIME - INCR+|DECR-
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:DATA?
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:DATA?
Description
Gets the data from buffer1 Contact1. Maximum buffer size is 1000. (Query Only)
NOTE
The transfer format is based on the setting of
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:FEED:CONTrol:INTerval on
page 250
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:FEED:CONTrol[:STATe] on page 251
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:POINts on page 252
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
1, 2, 4, 6, 8
Specifies measurement speed (time).