:INITiate:CONTinuous
Syntax
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description
This command changes the trigger state from “idle” state to the “wait for trigger”
automatically or stay in “idle” state. After executing
, this value changes
to ON, after
*RST
this value changes to OFF .
For details on the trigger system, refer to
NOTE
This command is turned ON when the trigger source is changed from the front panel of the
E4981A.
Parameters
This setup is initialized to ON when using the
command and to OFF
when using the
Query response
{1|0}<newline><^END>
Related commands
*RST on page 167
:SYSTem:PRESet on page 279
Equivalent key
sequence
No equivalent key is available on the front panel.
:INITiate[:IMMediate]
Syntax
:INITiate[:IMMediate]
Description
Changes the trigger state from “idle” state to “wait for trigger” state for one trigger
sequence. (No Query)
For details on the trigger system, refer to
. (No query)
Related commands
:INITiate:CONTinuous on page 224
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1
Enables successive startup.
OFF or 0 (initial value)
Disables successive startup.