276
Chapter 10
Command Reference
E4981A SCPI Commands
:SYSTem:FSHift
Syntax
:SYSTem:FSHift <numeric>
:SYSTem:FSHift?
Description
When you make a measurement with the measurement frequency set to 1 MHz, specifies
the shift of the signal frequency (frequency shift value) actually applied to the DUT
relative to 1 MHz as a percentage of 1 MHz.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value or the
maximum value is set.
NOTE
A fraction below the resolution is rounded off. MAX or MIN can be used to specify the
parameters.
This command is not available in Option 002.
The value of this command is not changed with
and is only
changed with Factory Default reset (available only through Front Panel).
Query response
{numeric}<newline><^END>
Equivalent key
sequence
[
Meas Setup
] - FREQ SHFT- 0%|1%|-1%|2%|-2%
<numeric>
Description
The value of the frequency shift.
Range
-2 to 2
Initial (Factory
set) value
0
Unit
% (percent)
Resolution
1