[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]
Syntax
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]?
Description
Determines whether to define the standard values for LOAD correction for each channel
individually when using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
command).
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:CKIT:STAN3[:DATA] on page 231
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - LOAD REF - MULTI|SINGLE
Description
ON or 1
Enables channel-by-channel value definition.
OFF or 0
(initial value)
Disables channel-by-channel value definition (defines the same