Chapter 10
175
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:CALCulate1:COMParator:AUXBin
Syntax
:CALCulate1:COMParator:AUXBin {ON|OFF|1|0}
:CALCulate1:COMParator:AUXBin?
Description
Turns ON/OFF the AUX_BIN function for sorting of the comparator function.
Depending on the ON/OFF state of the AUX_BIN function, the following difference
occurs in the sorting result when the measurement result of the secondary parameter
exceeds the limit range.
ON:
Sorted into AUX_BIN if the measurement result of the primary
parameter is within the limit range. Otherwise, sorted into
OUT_OF_BINS.
OFF:
Always sorted into OUT_OF_BINS.
Parameters
Query response
{1|0}<newline><^END>
Equivalent key
sequence
[
Meas Setup
] - LIMIT TABLE - AUX
:CALCulate1:COMParator:BEEPer:CONDition
Syntax
:CALCulate1:COMParator:BEEPer:CONDition {FAIL|PASS}
:CALCulate1:COMParator:BEEPer:CONDition?
Description
Sets the condition for producing a beep sound: when sorting with the comparator fails
(sorts into any BIN other than BIN1 to BIN9) or passes (sorts into BIN1 to BIN9).
Parameters
Query response
{FAIL|PASS}<newline><^END>
Related commands
:SYSTem:BEEPer:STATe on page 269
:CALCulate1:COMParator:BEEPer[:STATe] on page 176
Equivalent key
sequence
[
Meas Setup
] - LIMIT TABLE - BEEP
Description
ON or 1
Turns ON the AUX_BIN function.
OFF or 0 (initial value)
Turns OFF the AUX_BIN function.
Description
FAIL (initial value) Instructs the instrument to beep when the sorting fails.
PASS
Instructs the instrument to beep when the sorting pass.