366
Appendix E
Error Messages
Error messages (order of error number)
Error messages (order of error number)
Error messages are displayed in the lower row of the E4981A’s display. You can read them
out by using the SCPI command. This section provides a description of each error message
and its remedy.
NOTE
Errors with a negative error number are basically general errors defined by IEEE488.2 for
GPIB/LAN/USB instruments. On the other hand, errors with a positive error number are
defined specifically for the E4981A.
16
Reference Measurement Aborted
This error occurs when REF data measurement is aborted.
21
1 MHz opt. not installed
This error occurs when cable correction commands are sent to E4981A with Option 002,
via GPIB/LAN/USB. This error is not generated with front panel operation.
41
Correction Measurement Aborted
This error occurs when the correction data measurement is aborted.
43
Measurement failed
A measurement failure has occurred during measuring the correction data.
46
LOAD measurement incomplete
This error occurs when the cable correction LOAD measurement is incomplete.
47
OPEN measurement incomplete
This error occurs when the cable correction OPEN measurement is incomplete.
48
REF measurement incomplete
This error occurs when the cable correction REF measurement is incomplete.
82
Store failed
This error occurs when external mass storage device fails or internal FLASH ROM
hardware fails.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
83
No data to load
There is no setup data for the selected number or no external mass storage device is
connected.
1070
Fan failed
Cooling fan hardware failure is detected.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.