Chapter 6
75
Reading Out Measured Result
Reading Out Measured Result
6.
Re
ad
in
g
ou
t Me
as
ur
ed
Result
Figure 6-9
Structure of data buffer
Procedure for use
Step 1.
Use the
:DATA:POINts:BUF1 command on page 200
to set up the number of
measurements you want to feed into the data buffer.
Step 2.
When you use buffer 1 or buffer 2, use the
:DATA:FEED:BUF1 command on page 194
to select whether the parameter is fed from the primary or secondary parameter.
Step 3.
Use the
:DATA:FEED:CONTrol[:STATe] command on page 198
to make the setup
feed the measured result into the data buffer.
Step 4.
Execute the measurement for the number of times specified in Step 1.
Step 5.
Use the
:DATA[:DATA] command on page 206
to read out the data temporarily stored in
the data buffer.
Step 6.
Repeat Steps 4 to 5.
Sample program
See
to view an example of reading out measured result using data buffer.