Chapter 10
263
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:SOURce:FREQuency[:CW]
Syntax
:SOURce:FREQuency[:CW] <numeric> [Hz|kHz|MHz]
:SOURce:FREQuency[:CW]?
Description
Sets the measurement frequency.
Depending on this setting, the available measurement range varies (set with the
[:SENSe][:FIMPedance]:RANGe[:UPPer]
command). If changing the measurement
frequency results in a conflict with the measurement range setting, an acceptable range is
automatically selected.
Parameters
Query response
{120|1E3|1E6}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:RANGe[:UPPer] on page 262
Equivalent key
sequence
[
Meas Setup
] - FREQ - 120 Hz|1 kHz|1 MHz
:SOURce:VOLTage:ALC[:STATe]
Syntax
:SOURce:VOLTage:ALC[:STATe] {ON|OFF|1|0}
:SOURce:VOLTage:ALC[:STATe]?
Description
Determines whether to use the signal level compensation function (LVL COMP).
Parameters
Query Response
{1|0}<newline><^END>
Equivalent key
sequence
[
Meas Setup
] - LVL COMP- ON|OFF
Description
Range
Option 001 = 120 Hz, 1kHz and 1MHz
Option 002 = 120 Hz and 1 kHz
Initial value
1 kHz
Description
ON or 1
Enables the signal level compensation function.
OFF or 0 (Initial value)
Disables the signal level compensation function.